Multi 9310 What to do if...
ba75937e03 12/2011 91
Display
Instrument does not
react to keystroke
You want to know
which software
version is in the
meter or IDS sensor
Cause Remedy
Batteries almost empty Replace the batteries (see section
13.1 M
AINTENANCE, page 86)
Cause Remedy
Operating condition undefined
or EMC load unallowed
Processor reset:
Press the <ENTER> and <On/Off>
key simultaneously
Cause Remedy
E. g., a question by the service
department
Switch on the meter.
Open the menu, <ENTER__> / Stor-
age & config / System / Service infor-
mation. The instrument data are
displayed.
or
Connect the sensor.
Press softkey [<F1>Info<F1>] /
[More] The sensor data are displayed
(see section 4.1.5 S
ENSOR INFO, page
18)