EasyManuals Logo

wtw Multi 9310 User Manual

wtw Multi 9310
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
Multi 9310 pH value
ba75937e03 12/2011 35
Continuing with two-
point calibration
9. Set the nominal buffer value for the measured temperature with
<S
SS
S><T
TT
T>.
10. Accept the calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
11. If necessary, finish the calibration procedure as a single-point calibra-
tion with <M>.
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS
pH sensor.
01.07.2011
08:00
12. Thoroughly rinse the sensor with deionized water.
13. Immerse the pH sensor in buffer solution 2.
14. When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it with
<S
SS
S><T
TT
T>.
15. Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
16. Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<ENTER>.
The pH value of the buffer solution is displayed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the wtw Multi 9310 and is the answer not in the manual?

wtw Multi 9310 Specifications

General IconGeneral
Brandwtw
ModelMulti 9310
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals