Multi 9310 Conductivity
ba75937e03 12/2011 61
Calibration
evaluation
After calibration, the meter automatically evaluates the current status of the cal-
ibration. The evaluation appears on the display and in the calibration record.
Calibration record
(USB output)
Menu item Setting/
function
Description
Calibration /
Calibration data
storage / Display
- Displays the calibration records.
Further options:
z Scroll through the calibration
records with <S
SS
S><T
TT
T>.
z Output the displayed calibration
record to the interface with <F2>/
[USB output].
z Output all calibration records to the
interface with <F2__>[USB out-
put].
z Quit the display with <F1>/[Back]
or <ENTER>.
z Switch directly to the measured
value display with <M>.
Calibration /
Calibration data
storage / Output to
RS232/USB
- Outputs the calibration data storage
to the interface
Display Calibration record Cell constant [cm
-1
]
+++ Within the range
0.450 ... 0.500 cm
-1
Error Error Outside the range
0.450 ... 0.500 cm
-1
Error elimination (see section 14
W
HAT TO DO IF..., page 88)
CALIBRATION Cond
Calibration date 01.07.2011 07:43:33
TetraCon 925
Ser. no. 09250033
Cell constant 0.476 1/cm 25.0 °C
Sensor +++