Multi 9310 Settings
ba75937e03 12/2011 67
Calibration /Cali-
bration interval
1 ... 180 ...
999 d
Calibration interval for the D.O. sensor
(in days).
The meter reminds you to calibrate
regularly by the flashing sensor sym-
bol in the measuring screen.
Calibration /Com-
parison meas.
on
off
Enables to adjust the measured value
with the aid of a comparison measure-
ment, e.g. Winkler titration.
For details, see section 7.3 C
ALIBRA-
TION, page 51.
FDO Check /
Start FDO Check
- Starts the FDO
®
check procedure
FDO Check /
Check interval
1 ... 60 ...
999 d
Interval for the FDO Check (in days).
The meter reminds you to check the
sensor regularly by FDO Check status
indicator in the measuring screen.
Sal correction on
off
Manual salt content correction for con-
centration measurements.
Salinity 0.0 ... 70.0 Salinity or salinity equivalent for the
salt content correction.
This menu item is only available if the
manual salt content correction is
switched on.
Response time
t90
30 ... 300 s Response time of the signal filter (in
seconds).
A signal filter in the sensor reduces the
limits of variation of the measured
value. The signal filter is characterized
by the response time t90. This is the
time after which 90 % of a signal
change is displayed.
Reset - Resets all sensor settings to the deliv-
ery condition (see section 9.6.1
R
ESETTING THE MEASUREMENT SET-
TINGS, page 72).
Menu item Possible
setting
Description