<7. Trouble Shooting> 7-1
IM 80J01A01-01E
7 Trouble Shooting
7.1 Trouble Shooting during Measurement Preperation
Probable Cause / Corrective Action
The power cannot be turned ON.
It might be that the power plug is not
connected to the power outlet.
Check the connection between
them.
It might be that CQ1 Main Unit and
Utility Box are not conneted by the
power calbe.
Check the connection between
them.
It might be that the main power switch
on the front panel of Utility Box is not
turned ON.
Turn the Power switch ON.
The door of CQ1 Main Unit can not
open.
It might be that the initilalization does
not end.
Wait until the initialization ends.
It might be that the measurement is
being processed.
Wait until the measurement ends.
It may take time for door opening
in case that stage speed is set to
low.
It might be that the door of CQ1 Main
Unit is open.
Close it.
It might be that sub cover of CQ1
Main Unit is open.
Close it.
The micorplate can not be set on the
stage.
It might be that the stage attachment
for microplates is not set on the
stage.
Set the stage attachment on the
stage.
The microplate can not be set
depending on its shape. For detail,
please contact the dealer
1
.
The slideglass(es) can not be set on
the stage.
It might be that the stage attachment
is not set on the stage.
Set the stage attachment on the
stage.
It might be that the sample holder for
slideglass is not set on the stage
attachment.
Set the slideglass(es) on the
sample holder for slideglass and
set sample holder on the stage
attachment.