EasyManua.ls Logo

YOKOGAWA CQ1 - Starting and Shutting Down of CQ1

YOKOGAWA CQ1
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-8 <3. Name of Each Part/Starting and Shutting Down the Equipment >
IM 80J01A01-01E
3.2 Starting and Shutting Down of CQ1
3.2.1 Starting CQ1
(1) Turn the power switch ON, then turn the key switch ON..
CAUTION
Please confirm below before starting the equipment.
Is the equipment correctly installed?
(Refer to 2.2 Installing the Equipment)
Are the cables correctly connected?
(Refer to 2.3 Wiring the Peripherals)
Immediately pull out the AC cord from inlet of equipment, when
you notice smoke coming out of the main unit, foul smell,
abnormal noise or any other abnormality.
(2) Power ON the Workstation.
(3) Confirm that there is enough free space in data (D:) disk drive.
Data volume of one image that CQ1 outputs is about 11MB. Keep
free space in data (D:) disk drive by considering total image
quantity to acquire.
Also, free space of data (D:) disk drive is smaller than 1GB,
measurement setting cant be carried out normally.
Power Switch
Key Switch

Related product manuals