<7. Trouble Shooting> 7-5
IM 80J01A01-01E
Problem
Probable Cause / Corrective Action
The laser autofocus does not operate
properly.
Kind of sample vessel might be
wrong.
󳊜 Confirm that sample vessel
corresponds to that of designated
in measurement protocol.
It might be that sample vessel which
is not registered is used.
󳊜 Register the sample vessel (refer
to page 5-19)
󳊜 Use registered sample vessel
Is gel used for cell culture?
󳊜 In Sample Vessel Type of
Sample window, check Laser AF
Bottom Detector. (Refer to page
5-19)
Is sample vessel whose cell culture
surface is processed especially?
󳊜 In Sample Vessel Type of
Sample window, check Laser AF
Bottom Detector. (Refer to page
5-19)
󳊜 In case of measuring other sample
in same protocol again, please
load the protocol in each time.
The Image autofocus does not
operate properly.
It might be that sample is too dark or
bright dusts exist in sample vessel.
󳊜 Uncheck ImageAF
Quality of map image is bad
󳊜 Zoom map image, and map image
is shown in original resolution
Despite focusing by Search
function, images acquired by Rec
are defocused.
In case that Rec position is far from
Search position, Z position that
sample exist may different.
󳊜 Perform focusing by Search at
Rec position.
Waiting occurs during the
measurement.
It might occur in case when image
acquisition and heavy duty analysis
are processed concurrently.
󳊜 It is recommended that image
acquisition and analysis are
processed separately.
Rec is unavailable
It might be that imaging Area is not
set.
󳊜 Set imaging Area.
It might be that well for Rec isn’t
selected.
󳊜 Select well for Rec.
The error occurs just after Rec
button being clicked.
It might be that the parameter(s) set
in analysis is (are) not appropriate.
󳊜 Reset the parameters.
It might occur that memories of
workstation run out because of heavy
duty analysis.
󳊜 Reset protocol.
It might occur that memories of
workstation run out because of too
many number of Z stack.
󳊜 Reduce the number of Z stack.