EasyManuals Logo

YOKOGAWA DL9040 User Manual

YOKOGAWA DL9040
256 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #91 background imageLoading...
Page #91 background image
5-57
IM 701310-17E
Commands
1
2
3
4
5
6
App
Index
:ANALysis:WPARameter<x>:LIST:VALue?
Function Queries the automated measured value of the
list display number of the waveform parameter
measurement.
Syntax
:ANALysis:WPARameter<x>:LIST:
VALue? {<NRf>|MAXimum}
<x> = 1 or 2
<NRf> = 1 to 100000
Example
:ANALYSIS:WPARAMETER1:LIST:VALUE? 1
-> :ANALYSIS:WPARAMETER1:LIST:
VALUE 1,1.000E+00
Description If the measurement is not possible, “NAN (Not A
Number” is returned.
If the value of the list display number is not present,
“NAN (Not A Number)” is returned.
If MAXimum is specified, the maximum list display
number is selected.
:ANALysis:WPARameter<x>:MODE
Function Sets the mode of the waveform parameter
measurement or queries the current setting.
Syntax
:ANALysis:WPARameter<x>:MODE
{HISTogram|LIST|TRENd}
:ANALysis:WPARameter<x>:MODE?
<x> = 1 or 2
Example
:ANALYSIS:WPARAMETER1:MODE HISTOGRAM
:ANALYSIS:WPARAMETER1:MODE?
-> :ANALYSIS:WPARAMETER1:MODE HISTOGRAM
:ANALysis:WPARameter<x>:TRACe<x>?
Function Queries all settings related to the trace of the
waveform parameter measurement.
Syntax
:ANALysis:WPARameter<x>:TRACe<x>?
<x> of WPARameter<x> = 1 or 2
<x> of TRACe<x> = 1 to 8
Example
:ANALYSIS:WPARAMETER1:TRACE1?
-> :ANALYSIS:WPARAMETER1:TRACE1:AREA1:
TYPE BURST
:ANALysis:WPARameter<x>:TRACe<x>:
AREA<x>?
Function Queries all settings related to the area of the
waveform parameter measurement.
Syntax
:ANALysis:WPARameter<x>:TRACe<x>:
AREA<x>?
<x> of WPARameter<x> = 1 or 2
<x> of TRACe<x> = 1 to 8
<x> of AREA<x> = 1 or 2
Example
:ANALYSIS:WPARAMETER1:TRACE1:AREA1?
-> :ANALYSIS:WPARAMETER1:TRACE1:AREA1:
TYPE BURST
:ANALysis:WPARameter<x>:TRACe<x>:
AREA<x>:TYPE
Function Sets the waveform parameter of the waveform
parameter measurement or queries the current
setting.
Syntax
:ANALysis:WPARameter<x>:TRACe<x>:
AREA<x>:TYPE {<Parameter>}
:ANALysis:WPARameter<x>:TRACe<x>:
AREA<x>:TYPE?
<x> of WPARameter<x> = 1 or 2
<x> of TRACe<x> = 1 to 8
<x> of AREA<x> = 1 or 2
<Parameter> = {BURSt|CMEan|COUNt|CRMS|
CSDeviation|DELay|DT|DUTYcycle|FALL|
FREQuency|HIGH|HILow|LOW|MAXimum|MEAN|
MINimum|NOVershoot|NWIDth|PERFrequency|
PERiod|POVershoot|PTOPeak|PWIDth|RISE|
RMS|SDEViation|TYCInteg|TYINteg|V1|V2}
Example
:ANALYSIS:WPARAMETER1:TRACE1:AREA1:
TYPE BURST
:ANALYSIS:WPARAMETER1:TRACE1:AREA1:
TYPE? -> :ANALYSIS:WPARAMETER1:TRACE1:
AREA1:TYPE BURST
:ANALysis:WPARameter<x>:TRENd?
Function Queries all settings related to the trend display of the
waveform parameter measurement.
Syntax
:ANALysis:WPARameter<x>:TRENd?
<x> = 1 or 2
Example
:ANALYSIS:WPARAMETER1:TREND?
-> :ANALYSIS:WPARAMETER1:TREND:
HRANGE MAIN;HSPAN 1;MEASURE:CURSOR:C1:
POSITION 1.000E+00;STATE 1;:ANALYSIS:
WPARAMETER1:TREND:MEASURE:CURSOR:C2:
POSITION 2.000E+00;STATE 1;:ANALYSIS:
WPARAMETER1:TREND:MEASURE:CURSOR:DC:
STATE 1;:ANALYSIS:WPARAMETER1:TREND:
MEASURE:MODE CURSOR;:ANALYSIS:
WPARAMETER1:TREND:VERTICAL 2.000E+00,
1.000E+00
5.3 ANALysis Group

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL9040 and is the answer not in the manual?

YOKOGAWA DL9040 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL9040
CategoryTest Equipment
LanguageEnglish

Related product manuals