5-142
IM DLM4038-17EN
:MEASure:{CHANnel<x>|MATH<x>}:DPRoxi
mal:MODE
Function Sets or queries the distal, mesial, and proximal
point mode setting.
Syntax
:MEASure:{CHANnel<x>|MATH<x>}:DPRoxi
mal:MODE {PERCent|UNIT}
:MEASure:{CHANnel<x>|MATH<x>}:DPRoxi
mal:MODE?
<x> of CHANnel<x> = 1 to 8
<x> of MATH<x> = 1 to 4
Example
:MEASURE:CHANNEL1:DPROXIMAL:
MODE PERCENT
:MEASURE:CHANNEL1:DPROXIMAL:MODE?
-> :MEASURE:CHANNEL1:DPROXIMAL:
MODE PERCENT
:MEASure:{CHANnel<x>|MATH<x>}:DPRoxi
mal:PERCent
Function Sets or queries the distal, mesial, and proximal
points as percentages.
Syntax
:MEASure:{CHANnel<x>|MATH<x>}:DPRoxi
mal:PERCent {<NRf>,<NRf>,<NRf>}
:MEASure:{CHANnel<x>|MATH<x>}:DPRoxi
mal:PERCent?
<x> of CHANnel<x> = 1 to 8
<x> of MATH<x> = 1 to 4
<NRf> = 0 to 100% (in 1% steps)
Example
:MEASURE:CHANNEL1:DPROXIMAL:
PERCENT 40,60,80
:MEASURE:CHANNEL1:DPROXIMAL:
PERCENT?
-> :MEASURE:CHANNEL1:DPROXIMAL:PERCE
NT 40,60,80
:MEASure:{CHANnel<x>|MATH<x>}:DPRoxi
mal:UNIT
Function Sets or queries the distal, mesial, and proximal
points as voltages.
Syntax
:MEASure:{CHANnel<x>|MATH<x>}:DPRoxi
mal:UNIT {<Voltage>,<Voltage>,<Volta
ge>|<NRf>,<NRf>,<NRf>}
:MEASure:{CHANnel<x>|MATH<x>}:DPRoxi
mal:UNIT?
<x> of CHANnel<x> = 1 to 8
<x> of MATH<x> = 1 to 4
<Voltage>, <NRf> = See the DLM4000 Features
Guide for this information.
Example
:MEASURE:CHANNEL1:DPROXIMAL:
UNIT−50V,0V,50V
:MEASURE:CHANNEL1:DPROXIMAL:UNIT?
-> :MEASURE:CHANNEL1:DPROXIMAL:
UNIT−50.0E+00,0.0E+00,50.0E+00
Description When the probe type is set to current, this
command sets or queries the <Current> value.
:MEASure:{CHANnel<x>|MATH<x>}:METHod
Function Sets or queries the high/low point setting method.
Syntax
:MEASure:{CHANnel<x>|MATH<x>}:METH
od {AUTO|MAXimum|HISTogram}
:MEASure:{CHANnel<x>|MATH<x>}:METH
od?
<x> of CHANnel<x> = 1 to 8
<x> of MATH<x> = 1 to 4
Example
:MEASURE:CHANNEL1:METHOD AUTO
:MEASURE:CHANNEL1:METHOD?
-> :MEASURE:CHANNEL1:METHOD AUTO
:MEASure:CONTinuous?
Function Queries all settings for the normal statistical
processing of automatically measured waveform
parameters.
Syntax
:MEASure:CONTinuous?
:MEASure:CONTinuous:RESTart
Function Restarts the normal statistical processing of
automatically measured waveform parameters.
Syntax
:MEASure:CONTinuous:RESTart
Example
:MEASURE:CONTINUOUS:RESTART
:MEASure:CONTinuous:TLCHange (Trigger
Level Change)
Function Sets or queries whether normal statistical
processing of automatically measured waveform
parameters is restarted when the trigger level is
changed.
Syntax
:MEASure:CONTinuous:TLCHange {RESTar
t|IGNore}
:MEASure:CONTinuous:TLCHange?
Example
:MEASURE:CONTINUOUS:TLCHANGE RESTART
:MEASURE:CONTINUOUS:TLCHANGE?
-> :MEASURE:CONTINUOUS:TLCHANGE REST
ART
:MEASure:CYCLe?
Function Queries all cyclic statistical processing settings
for automated measurement.
Syntax
:MEASure:CYCLe?
:MEASure:CYCLe:ABORt
Function Aborts cyclic statistical processing for automated
measurement.
Syntax
:MEASure:CYCLe:ABORt
Example
:MEASURE:CYCLE:ABORT
:MEASure:CYCLe:EXECute
Function Executes cyclic statistical processing for
automated measurement.
Syntax
:MEASure:CYCLe:EXECute
Example
:MEASURE:CYCLE:EXECUTE
5.21 MEASure Group

Table of Contents

Other manuals for YOKOGAWA DLM4038

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM4038 and is the answer not in the manual?

YOKOGAWA DLM4038 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM4038
CategoryTest Equipment
LanguageEnglish

Related product manuals