5-228
IM DLM4038-17EN
:SERialbus<x>:SPI[:ANALyze]:SETup:FI
ELd
Function Sets or queries the data field size for SPI bus
signal analysis.
Syntax
:SERialbus<x>:SPI[:ANALyze]:SETup:FI
ELd {<NRf>}
:SERialbus<x>:SPI[:ANALyze]:SETup:FI
ELd?
<x> = 1 to 4
<NRf> = 4 to 32
Example
:SERIALBUS1:SPI:ANALYZE:SETUP:FIE
LD 4
:SERIALBUS1:SPI:ANALYZE:SETUP:FIELD?
-> :SERIALBUS1:SPI:ANALYZE:SETUP:FIE
LD 4
:SERialbus<x>:SPI[:ANALyze]:SETup:GR
Ouping
Function Sets or queries the on/off status of grouping for
SPI bus signal analysis when there is no chip
select signal.
Syntax
:SERialbus<x>:SPI[:ANALyze]:SETup:GR
Ouping {<Boolean>}
:SERialbus<x>:SPI[:ANALyze]:SETup:GR
Ouping?
<x> = 1 to 4
Example
:SERIALBUS1:SPI:ANALYZE:SETUP:GROUPI
NG ON
:SERIALBUS1:SPI:ANALYZE:SETUP:GROUPI
NG?
-> :SERIALBUS1:SPI:ANALYZE:SETUP:GRO
UPING 1
:SERialbus<x>:SPI[:ANALyze]:SETup:IT
IMe
Function Sets or queries the idle time for SPI bus signal
analysis when there is no chip select signal.
Syntax
:SERialbus<x>:SPI[:ANALyze]:SETup:IT
IMe {<Time>}
:SERialbus<x>:SPI[:ANALyze]:SETup:IT
IMe?
<x> = 1 to 4
<Time> = 10 ns to 1 ms
Example
:SERIALBUS1:SPI:ANALYZE:SETUP:ITI
ME 2US
:SERIALBUS1:SPI:ANALYZE:SETUP:ITIME?
-> :SERIALBUS1:SPI:ANALYZE:SETUP:ITI
ME2.00000E−06
:SERialbus<x>:SPI[:ANALyze]:SETup:MO
DE
Function Sets or queries the data signal wiring system
(three-wire or four-wire) for SPI bus signal
analysis.
Syntax
:SERialbus<x>:SPI[:ANALyze]:SETup:MO
DE {WIRE3|WIRE4}
:SERialbus<x>:SPI[:ANALyze]:SETup:MO
DE?
<x> = 1 to 4
Example
:SERIALBUS1:SPI:ANALYZE:SETUP:MO
DE WIRE3
:SERIALBUS1:SPI:ANALYZE:SETUP:MODE?
-> :SERIALBUS1:SPI:ANALYZE:SETUP:MO
DE WIRE3
:SERialbus<x>:SPI[:ANALyze]:SETup:MS
BLsb
Function Sets or queries the data MSB and LSB bits for
SPI bus signal analysis.
Syntax
:SERialbus<x>:SPI[:ANALyze]:SETup:MS
BLsb {<NRf>,<NRf>}
:SERialbus<x>:SPI[:ANALyze]:SETup:MS
BLsb?
<x> = 1 to 4
<NRf> = 0 to 31
Example
:SERIALBUS1:SPI:ANALYZE:SETUP:MSBL
SB 7,0
:SERIALBUS1:SPI:ANALYZE:SETUP:MSBL
SB?
-> :SERIALBUS1:SPI:ANALYZE:SETUP:MSB
LSB 7,0
:SERialbus<x>:SPI:DETail?
Function Queries all SPI bus signal analysis result list
settings.
Syntax
:SERialbus<x>:SPI:DETail?
<x> = 1 to 4
:SERialbus<x>:SPI:DETail:DISPlay
Function Sets or queries the display mode for the SPI bus
signal analysis result list.
Syntax
:SERialbus<x>:SPI:DETail:DISPl
ay {FULL|LOWer|UPPer}
:SERialbus<x>:SPI:DETail:DISPlay?
<x> = 1 to 4
Example
:SERIALBUS1:SPI:DETAIL:DISPLAY FULL
:SERIALBUS1:SPI:DETAIL:DISPLAY?
-> :SERIALBUS1:SPI:DETAIL:DISPLAY FU
LL
5.25 SERialbus Group

Table of Contents

Other manuals for YOKOGAWA DLM4038

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM4038 and is the answer not in the manual?

YOKOGAWA DLM4038 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM4038
CategoryTest Equipment
LanguageEnglish

Related product manuals