Axiotech and Axiotech
vario
Carl Zeiss
B 40-020 e 06/99 1-19
1.11
Technical Data
Dimensions (width × depth × height)
Axiotech stand...........................................................................................approx. 210 × 315 × 435 mm
Axiotech
vario
stand.......................................................................................approx. 412 × 452 × 375 mm
Axiotech footprint................................................................................................approx. 250 × 350 mm
Axiotech
vario
footprint............................................................................................approx. 420 × 460 mm
Weight
Axiotech............................................................................................................................approx. 12 kg
Axiotech
vario
........................................................................................................................approx. 22 kg
Ambient conditions
Storage and transport (in packaging):
Permissible ambient temperature .......................................................................................-40 to +50 °C
Permissible relative humidity (without condensation)............................................................... max. 80 %
Atmospheric pressure .............................................................................................800 hPa to 1060 hPa
Operation:
Permissible ambient temperature ........................................................................................+5 to +35 °C
Permissible relative humidity (without condensation)............................................................... max. 80 %
Altitude for application .......................................................................................................max. 2000 m
Atmospheric pressure .............................................................................................800 hPa to 1060 hPa
Pollution degree....................................................................................................................................2
Operating data
Areas of use .......................................................................................................................closed rooms
Protection class......................................................................................................................................I
Protection type................................................................................................................................IP 20
Electrical safety.......................................................................according to IEC 1010-1 (DIN EN 61010-1)
.........................................................................................taking CSA and UL specifications into account
Excess voltage category......................................................................................................................... II
Radio interference suppression..................................................................according to EN 55011 Class B
Line voltage ranges:
-
wide-range power unit 6 V 30 W, stabilized...................................................100 to 240 V AC ±10 %
-
integrated power unit 12 V 100 W, stabilized.................100 to 127 V AC or 220 to 240 V AC ±10 %
-
external power supply 6 V 10 ... 40 W, 2 ports, switchable .......................................120 V AC ±10 %
........................................................................................................................230 V AC + 6 / - 10 %
-
Power unit for HBO 50..................................................................................220 to 240 V AC ±10 %
-
Transformer N XBO 75..................................................................................100 to 240 V AC ±10 %
-
Separate transformer SNT 12 V DC 100 W......................100 to 127 V AC or 220 to 240 V AC ±10 %