EasyManua.ls Logo

Zeiss Axiotech - Attachment, Change and Centering of the Microscope Stage; Axiotech 30 and Axiotech 100 with Stage Carrier a

Zeiss Axiotech
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Carl Zeiss Axiotech and Axiotech
vario
2-12 B 40-020 e 06/99
• Fix condenser in position with the clamping screw (2-13/4) and move it close to the uppermost stop
position via the condenser drive (2-13/2).
☞
Section 1.6 contains an overview of condensers available for the Axiotech.
The maximum specimen height for use of stage carrier D is 39 mm.
2.8
Attachment, change and centering of the microscope stage
When delivered, the microscope is factory-aligned in accordance with the order. The possibly required
change of the stage between reflected-light and transmitted-light applications is described below.
2.8.1
Axiotech 30 and Axiotech 100 with
stage carrier A
• Slightly tilt the microscope stand backwards.
• Loosen four hexagonal screws (2-14/1) on the
stage carrier A using the SW3 ball-headed
screwdriver (2-14/2).
• Remove microscope stage (2-14/3) and replace
it with another stage.
• Screw the suitable specimen holder on the
microscope stage.
☞
Stage carrier A (2-14/4) remains attached. For specimens which are higher than 25 mm, the
stage carrier can be lowered after loosening the two screws (2-14/5). The maximum lowering
range is 24.5 mm, i.e. the specimen area is increased to the maximum specimen height of
49.5 mm.
Fig. 2-14 Attachment/change of microscope
stage (configuration with stage
carrier A)

Table of Contents