Pannoramic MIDI 1.17 Users Guide 5 Troubleshooting and Maintenance
8. Switch on the control computer.
9. Launch control software.
Illumination error! Probably the lamp has burned out. Please turn off the scanner and check the
lamp!
Halogen lamp is defective
Replace the halogen lamp (see section 5.3.2 Replacing the halogen lamp)
Image is black
Halogen lamp is defective
Replace halogen lamp (see section 5.3.2 Replacing the halogen lamp)
Virtual slide is improperly focused almost across its entire surface
Focus range enabled
Disable Limiting focus range function
Poor image quality
Imaging error (for example, chromatic aberrations)
Contact 3DHISTECH Service for support.
Warning!
Do not modify the camera driver settings, as the default factory
settings give the most accurate images.
Virtual slides show poor stitching in Pannoramic Viewer
Scan camera is not properly adjusted (for example, due to loose fixture)
Contact 3DHISTECH Service for support.
April 3, 2014 - Rev. 1 3DHISTECH Ltd.
110(127)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the 3D Histech Pannoramic MIDI and is the answer not in the manual?

3D Histech Pannoramic MIDI Specifications

General IconGeneral
Brand3D Histech
ModelPannoramic MIDI
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals