Pannoramic MIDI 1.18 Users Guide 4 Software Description
Figure 39 – Focus level and step size
With the Z-stack method separate layers are created and stored, and can be opened with
Pannoramic Viewer. The Z-stack slide contains the extended focus layer. Both for Extended
focus and Z-Stack scanning methods, you can determine the desired number of focus levels
with the Focus levels selection box. The current focus value is the starting point. If you set 5
for example, one current, two focus levels above and below FOVs are captured. Therefore, it
is advised to choose an odd number for the focus levels.
Focus step size determines the spacing between focus levels. One step is around 0.2 µm.
Default settings are: Number of focus levels: 7; Focus step size: 5.
4.3.6 Service panel
Figure 40 – Service panel
This panel contains control and maintenance functions.
Emergency Manual Slide Removal Position
The specimen holder will be moved to a position where the slide can be removed manually
after opening the cover (see section 5.3.1 Opening the protective covers).
Slide Removed Manually
Click this button to notify the software that you have removed the slide manually.
April 3, 2014 - Rev 1 3DHISTECH Ltd. 60(127)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the 3D Histech Pannoramic MIDI and is the answer not in the manual?

3D Histech Pannoramic MIDI Specifications

General IconGeneral
Brand3D Histech
ModelPannoramic MIDI
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals