Pannoramic MIDI 1.17 Users Guide 5 Troubleshooting and Maintenance
5.2 Troubleshooting
This section describes the most common troubleshooting solutions for the Pannoramic MIDI product.
If any problem arises that does not have solution in this manual, contact the service personnel.
The product does not start
Power supply is interrupted
1. Make sure that the 24V external power supply is on.
2. Check cable connections.
Control software has stopped
Slide jammed
1. Exit control software.
2. Switch off control computer and 24V power supply, disconnect power cable.
3. Remove protective covers. Follow the steps in section 5.3.1 Opening the protective
covers.
4. Remove slide manually (see section 5.3.4 Removing a slide).
5. Reattach protective covers.
6. Reconnect 24V power supply and control computer to line power supply.
7. Switch on the 24 V power supply.
8. Switch on the control computer.
9. Launch control software.
Tray jammed
1. Exit control software.
2. Switch off control computer and 24V power supply, disconnect power cable.
3. Remove protective covers (see section 5.3.1 Opening the protective cover).
4. Remove tray manually. If a jammed tray cannot be removed, contact service
personnel.
5. Reattach protective covers.
6. Reconnect 24V power supply and control computer to line power supply.
7. Switch on the 24 V power supply.
April 3, 2014 - Rev. 1 3DHISTECH Ltd.
109(127)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the 3D Histech Pannoramic MIDI and is the answer not in the manual?

3D Histech Pannoramic MIDI Specifications

General IconGeneral
Brand3D Histech
ModelPannoramic MIDI
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals