Pannoramic MIDI 1.17 Users Guide 5 Troubleshooting and Maintenance
5.3.5 Installing and Removing the Filter Module (Fluorescent Option)
Figure 82 – Installing and removing the filter module
1. If using an HXP 120 illumination unit, turn it off.
Warning!
Crush hazard! Stop the running digitization process before you proceed
with filter module installation or removal.
2. Pull the pair of the protective cover halves apart or remove the right part completely (see
section 5.3.1 Opening the protective covers), but do not switch the main unit off.
3. Select the desired filter in the corresponding software window (see section 4.2.3
Microscope Settings / Filter Settings). The filter will be positioned in such a way that the
user can access it without having to turn the filter wheel manually.
April 3, 2014 - Rev. 1 3DHISTECH Ltd.