EasyManuals Logo

3D Histech Pannoramic MIDI User Manual

Default Icon
127 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #51 background imageLoading...
Page #51 background image
Pannoramic MIDI 1.18 Users Guide 4 Software Description
General threshold options
Remove coverslip If selected, the edges of the coverslip are not detected as specimen to be
scanned. Depending on the threshold level, it is possible that the edges of the coverslip are
detected as areas to be digitized.
Figure 28 – Preview image with and without coverslip edge
Caution!
It requires lots of time (since it cannot find focus neither on the
edges, nor on specks) to digitize coverslip edges. Select Remove
coverslip function to eliminate those areas.
Warning!
Do not use Remove coverslip function on TMA slides! The layout of
TMA slides may mislead the detecting algorithm.
Fill holes in scan area If selected, the holes inside the specimen will also be scanned.
Often these holes contain small sample objects that otherwise would not be scanned.
Remove specks If selected, the scanner does not digitize objects smaller than the
predetermined size. This helps filtering specks and other uninteresting objects out. Set the
maximum size of specks not to be digitized by adjusting the Speck size slider.
Scan area dilationAdjust the slider To enlarge or decrease the scanned area. The edge is
extended discretely by 100 µm each time.
April 3, 2014 - Rev 1 3DHISTECH Ltd. 51(127)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the 3D Histech Pannoramic MIDI and is the answer not in the manual?

3D Histech Pannoramic MIDI Specifications

General IconGeneral
Brand3D Histech
ModelPannoramic MIDI
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals