18
Unit setting example
The example below sets the units in the order with g (gram) as the first unit followed by pc (counting
mode).
1 Press and hold the SAMPLE key until ba5fnc of the
function table is displayed, then release the key.
2 Press the SAMPLE key several times to display 1Unit .
3 Press the PRINT key to enter the unit selection mode.
4 Press the RE-ZERO key to specify the unit of g
The stabilization indicator appears when the unit
is specified.
5 Press the SAMPLE key to display 1Unit pC5 .
6 Press the RE-ZERO key to specify the unit of pc
The stabilization indicator appears when the unit
is specified.
7 Press the PRINT key to store the units.
The balance displays end and then displays the next
menu item of the function table.
8 Press the CAL key to exit the function table. Then the
balance returns to the weighing mode with g,the unit
selected first.
9 Press the MODE key to switch between g and pc (gpcs).
Press
several times
Select
Specify
Store
Displays the units
sequentially.