13-2 Statistical Calculation Mode (Example Of Use)
Here, as an example of use of the statistical calculation mode, mixing of
the multiple formulae such as medicine is described. The mixing process is
recorded using the balance and the printer.
In the example, the GX-303A and the AD-8126 or AD-8127 are connected
using the RS-232C serial interface.
(1) Getting Started
Changing the function table
Changes To enable the statistical calculation mode
To enable “Zero after output
Enabling the statistical calculation mode
1 Enter the function table menu.
Press and hold the SAMPLE key until ba5fnc of the
function table is displayed, then release the key.
2 Select the application function.
Press the SAMPLE key several times to display 1ap fnc .
Then, press the PRINT key to display 1 .
3 Change the application function parameter to “2”.
Press the RE-ZERO key to display
. Press the
PRINT key to confirm the change. After end ,
1 is displayed.
Enabling “Zero after output”
4 Select “Zero after output”.
Press the SAMPLE key several times to display 1dout . Then,
press the PRINT key to display , and press the
SAMPLE key several times to display 1 .
5 Enable “Zero after output”.
Press the RE-ZERO key to display
. Then, press
the PRINT key to confirm the change. After 1end , 5if is
displayed.
Returning to the weighing mode
6 Press the CAL key to return to the weighing mode.
Press several times
Press several times
Press several times
Returning to the weighing mode