21.Maintenance
21-1 Treatment Of The Balance
Clean the balance with a lint free cloth that is moistened with warm water and a mild detergent.
Do not use organic solvents to clean the balance.
Do not disassemble the balance.
Use the original packing material for transportation.
22.Troubleshooting
22-1 Checking The Balance Performance And Environment
The balance is a precision instrument. When the operating environment or the operating method is inadequate,
correct weighing can not be performed. Place a sample on the pan and remove it, and repeat this several
times. If the balance seems to have a problem with repeatability or to perform improperly, check as described
below. If improper performance persists after checking, contact the local A&D dealer for repair.
"Frequently asked questions" and their answers are also posted on our website
<http://www.aandd.co.jp>.
1Checking that the balance performs properly
Please check the operation of the balance by the self diagnosis function. Refer to "6-2 Self-diagnosis
Function". Fatal faults are indicated by messages.
Check the balance performance using an external weight. Be sure to place the weight in the center of the
weighing pan.
Check the balance repeatability, linearity and calibrated value using external weights with a known
value.
2.Checking that the operating environment or weighing method is proper
Operating environment
Is the weighing table solid enough?Especially 1mg model
Is the balance level? Refer to " 2-2 Precaution "How to adjust the bubble spirit level".
Is the operating environment free from vibration and drafts?
Is there a strong electrical or magnetic noise source such as a motor near the balance?
Weighing method
Does the weighing pan rim touch anything? Is the weighing pan assembly installed correctly?
Is the RE-ZERO key pressed before placing a sample on the weighing pan?
Is the sample placed in the center of the weighing pan?
Has the balance been calibrated using the internal mass (one-touch calibration)? (Only for GX=A
series)
Has the balance been warmed up for one hour before weighing?