Sample and container
Has the sample absorbed or lost moisture due to the ambient conditions such as temperature and
humidity?
Has the temperature of the container been allowed to equalize to the ambient temperature? Refer to
"2-3 During Use".
Is the sample charged with static electricity? Refer to "2-3 During Use".
(This occurs especially with 0.001g models when the relative humidity is low.)
Is the sample of magnetic material such as iron? There are cautions about weighing magnetic
materials. Refer to "2-3 During Use".