96
Display Error code Description
An undefined command was received.
Confirm the command.
EC,E02
Not ready
A received command can not be processed.
Example:
The balance received a "Q" command, but not in the
weighing mode.
The balance received a "Q" command while processing
a RE-ZERO command. Adjust the delay time to
transmit a command.
EC,E03
Timeout error
If the timeout parameter is set to "t-Up1", the balance did not
receive the next character of a command within the time limit
of one second. Confirm the communication.
EC,E04
Excess characters error
The balance received excessive characters in a command.
Confirm the command.
EC,E06
Format error
A command includes incorrect data.
Example:
The data is numerically incorrect.
Confirm the command.
EC,E07
Parameter setting error
The received data exceeds the range that the balance can
accept. Confirm the parameter range of the command.

Table of Contents