Parameterization
80 FSM4000 D184B140U02
Submenu/parameter Value range Comment
Submenu Set the time-out limits and alarm behavior for values over
or under the coil AC.
Alarm Coil Current
Min. value 0 ... 500 Lower range limit for alarm.
20.0 mA
Max. value 0 ... 500 Upper range limit for alarm.
500.0 mA
Message Off, warning, error Defines the alarm behavior.
Off
Submenu
Set the time-out limits and alarm behavior for values over
or under the coil resistance.
Note: To correctly display the coil resistance or coil
temperature, an adjustment is required or the cable length
must be entered. For additional information, see
"Submenu Adjustment" on page 85.
Alarm Coil Resist
Min. value 0 ... 1500 Lower range limit for alarm.
2 ohms
Max. value 0 ... 500 Upper range limit for alarm.
500 ohms
Message Off, warning, error Defines the alarm behavior.
Off Note: If coil resistance is being measured (at least every
60 s), the most recently acquired flowrate is retained
during the measurement (2 s approx.).
Submenu Set the time-out limits and alarm behavior for values over
or under the coil temperature.
Note: To correctly display the coil resistance or coil
temperature, an adjustment is required or the cable length
must be entered. For additional information, see
"Submenu Adjustment" on page 85.
Alarm Coil Temp
Min. value -100 ... 200 Lower range limit for alarm.
-50 °C
Max. value -100 ... 200 Upper range limit for alarm.
200 °C
Message Off, warning, error Defines the alarm behavior.
Off
Note: If coil resistance is being measured (at least every
60 s), the most recently acquired flowrate is retained
during the measurement (2 s approx.).