Parameterization
D184B140U02 FSM4000 81
Submenu/parameter Value range Comment
Submenu Set the time-out limits and alarm behavior for values over
or under the coil temperature.
Alarm Ins Resist
Min. value 0 ... 100000 Lower range limit for alarm.
1 megaohm
Max. value 0 ... 100000 Upper range limit for alarm.
50 megaohm
Message Off, warning, error Defines the alarm behavior.
Off
Note: If coil insulation resistance is being measured (at
least every 10 minutes), the most recently acquired
flowrate is retained during the measurement (2 s approx.).
Submenu
Set the time-out limits and alarm behavior for values over
or under the electrode AC voltage for electrode E1.
Alarm E1 Voltage
Min. value 0 ... 3000 Lower range limit for alarm.
0.0 µV
Max. value 0 ... 3000 Upper range limit for alarm.
3000.0 µV
Message Off, warning, error Defines the alarm behavior.
Off
Note: To properly display the electrode voltage, the
electrode zero must be adjusted. For additional
information, see "Submenu Adjustment" on page 85.
Submenu Set the time-out limits and alarm behavior for values over
or under the electrode AC voltage for electrode E2.
Alarm E2 Voltage
Min. value 0 ... 3000 Lower range limit for alarm.
0.0 µV
Max. value 0 ... 3000 Upper range limit for alarm.
3000.0 µV
Message Off, warning, error Defines the alarm behavior.
Off Note: To properly display the electrode voltage, the
electrode zero must be adjusted. For additional
information, see "Submenu Adjustment" on page 85.