EasyManua.ls Logo

Abbott Prodigy 3855 - Technical Support; Appendix A: Product Specifications; Patient Programmer Specifications

Abbott Prodigy 3855
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 8. Troubleshoong during postoperave programming
Problem Possible Cause Possible Soluon
Frequency is improperly set. Adjust sengs to a
comfortable level, avoiding
abrupt changes.
Acve electrodes are not posioned
over the correct target.
Set amplitude to zero and
program dierent
electrodes.
Paent has inadvertently changed the
programming.
Reprogram the system;
review operaon and usage
with the paent.
Painful Smulaon
Paent feels a burning sensaon. Lead is fractured or current is leaking. Xray the enre system for
lead fracture, breaks, or
sharp angles.
Paent has inadvertently changed the
programming.
Reprogram the system;
review operaon and usage
with the paent.
Excessive scarring or broc build-up is
present around the electrodes.
Order a new Xray and
compare it with the original.
Technical Support
For technical quesons and support for your product, use the following informaon:
+1
855
478
5833
(toll-free within North America)
+1 651 756 5833
For addional assistance, call your local Abbo Medical representave.
Appendix A: Product Specicaons
NOTE: Not all models are available in all countries. Contact your local representave for more informaon.
Paent Programmer Specicaons
The
Prodigy
paent programmer (Model 3855 or 3856) has the following physical specicaons.
Table 9. Paent programmer specicaons
MRI status MR Unsafe
Height 6.8 cm (2.7 in.)
Length 10.77 cm (4.2 in.)
Thickness (T) 2.6 cm (1.0 in.)
Weight 128 g (4.6 oz)
Expected service life 5 years
Storage temperature -10
°C–55°C (14°F–131°F)
Storage humidity 10
%–90% (noncondensing)
Storage pressure 70–150 kPa (10.2–21.8 psi)
Operang temperature 10
°C–40°C (50°F–104°F)
37