Abbott Prodigy 3855 - Appendix D: Electromagnetic Compatibility Guidelines

Abbott Prodigy 3855
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 12. Symbols and denions
Symbol Denion
which could be caused by inappropriate waste handling of this product. The
recycling of materials will help to conserve natural resources.
For more informaon about how to return this product for recycling, please
contact
Abbo
Medical
.
This device is listed by the Canadian Standards Associaon (CSA)
Internaonal as cered
Appendix D: Electromagnec Compability Guidelines
The
Prodigy
paent programmer, hereaer the device, is medical equipment and should be used with the
following guidance:
The device requires special precauons with regard to electromagnec compability (EMC) and should be
used in accordance with the informaon provided in this manual.
The device has no essenal performance, as dened by IEC 606011/AMD1:2012 (Edion 3.1). Performance of
the device was maintained during electromagnec disturbance tesng. The device is intended for use in the
electromagnec environment specied in the following tables. The user should ensure that it is used in such
an environment.
CAUTION: The device complies with the limits for medical devices contained in IEC 6060112:201404
and CISPR11:2015 (BS EN 55011:2016). However, the device may cause radio interference or may disrupt
the operaon of nearby equipment. It may be necessary to migate this eect by reorienng or
relocang the receiving device.
CAUTION: To avoid increasing emissions or decreasing immunity from a device or system, use only
components approved by
Abbo
Medical
with this system. Do not use
Abbo
Medical
components with
devices or systems that are not approved by
Abbo
Medical
.
Table 13. Guidance a nd Manufacturer's Declaraon
Electromagnec Emissions
Emissions Test Compliance Electromagnec Environment Guidance
RF emissions
CISPR 11
Group 1 The device produces RF emissions as a result of its internal and system interface funcons. Its
RF emissions are very low and are not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B The device is suitable for use in all establishments including domesc establishments and
those directly connected to the public low-voltage power supply network that supplies
buildings used for domesc purposes.
Harmonic emissions
IEC 6100032
Not applicable
Voltage uctuaons/icker
emissions IEC 6100033
Not applicable
NOTE: The RF output frequency of the device is 87.5 kHz. The eld strength is 32 dBμA/m at 3 m
(inducvely coupled device).
Table 14. Guidance and Manufacturer's Declaraon
Electromagnec Immunity
Immunity Test IEC 60601 Test Level Compliance Level Electromagnec
Environment
Guidance
Electrostac
discharge (ESD)
IEC 6100042
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
The
relave humidity should be at
least 30%.
Electrical fast transient/burst
IEC 6100044
±2 kV for power supply lines
±1 kV for input/output lines
Not applicable No guidance for
baery-powered
devices.
Surge
IEC 6100045
±1 kV line to line
±2 kV line to ground
Not applicable No guidance for
baery-powered
devices.
41