Chapter 2 Design the Relative Standard Curve Experiment
Set Up the Targets
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
28
Notes
Design
Guidelines
When you design your own relative standard curve experiment:
Select the Set Up Standards check box. Applied Biosystems recommends that you
set up a standard curve for each target assay in the reaction plate.
Identify each target assay with a unique name and color. You can enter up to
100 characters in the Target Name field.
Select an endogenous control for each sample. The endogenous control is a target
that is in all samples under investigation. It should be expressed equally in all
sample types, regardless of treatment or tissue origin (examples of endogenous
controls are β-actin, GAPDH, and 18S ribosomal RNA [18S rRNA]). The
endogenous control is used to normalize the PCR results; the endogenous control
corrects for variable sample mass, nucleic acid extraction efficiency, reverse
transcription efficiency, and pipette calibration errors. Note that:
Each sample type (for example, each tissue in a study comparing multiple tissues)
requires an endogenous control.
If samples are on multiple plates, each reaction plate must have an endogenous
control. Additionally, every plate must include an endogenous control for every
sample type on the reaction plate.
Select the reporter dye used in the target assay. In the Methods and Materials screen
on page 24, if you selected:
–TaqMan
®
Reagents – Select the dye attached to the 5 end of the TaqMan probe.
–SYBR
®
Green Reagents – Select SYBR.
Select the quencher used in the target assay. In the Methods and Materials screen on
page 24, if you selected:
–TaqMan
®
Reagents – Select the quencher attached to the 3 end of the TaqMan
probe.
–SYBR
®
Green Reagents – Select None.
For More
Information
For more information on:
Completing the Targets screen – Open the 7500 Software Help by clicking or
pressing F1.
Selecting an endogenous control – See the Application Note Using TaqMan
®
Endogenous Control Assays to Select an Endogenous Control for Experimental
Studies.