Chapter 2 Design the Relative Standard Curve Experiment
Review the Reaction Setup
Applied Biosystems 7500/7500 Fast Real-Time PCR System Getting Started Guide for Relative Standard Curve
and Comparative C
T
Experiments
40
Notes
To complete the Sample Dilution Calculations tab:
1. In the Reaction Setup screen, select the Sample Dilution Calculations tab.
2. Click the Diluted Sample Concentration (10✕ for Reaction Mix) field, then
enter 50.
3. In the unit drop-down list, select ng/µL (default).
4. Review the calculated volumes for the sample dilutions:
7b
7a
7c
Sample Name
Stock Concentration
(ng/µL)
Sample Volume
(µL)
Diluent Volume
(µL)
Total Volume of Diluted Sample
(µL)
Liver 100.0 17.00 17.00 34.00
Kidney 100.0 17.00 17.00 34.00
1
2
3
4