11-48 ADS TRITON+ Manual
Problem
Monitor independently discontinues logging data.
Possible
Causes
Battery pack may be dead or below minimum voltage
requirement
(7.5 volts - internal).
3.3
-volt (yellow) fuse on regulator may have blown.
Monitor memory may have been corrupted during
firmware upgrade, causing the monitor to record data
only
during monitor communications.
Possible
Solutions
Replace battery pack.
Replace the 3.3
-volt (yellow) fuse on the regulator.
If the monitor is logging data only during monitor
communications following a firmware upgrade, contact
your ADS regional representative to clear the monitor
memory and reactivate the monitor.
Contact your regional ADS representative.
Problem
An
I/O error message displays when communicating
with the monitor.
Possible
Causes
Communication quality might be marginal.
Antenna may be defective.
Processor board may be faulty.
Possible
Solutions
Re
-attempt communication with the monitor.
Temporarily attach new antenna, and permanently
replace the existing antenna if the error message no
longer occurs.
Replace the monitor if defective.
Contact your regional
ADS representative.