11-68 ADS TRITON+ Manual
Pressure Depth Subsystem
The following tables contain general troubleshooting techniques for
the pressure depth subsystem corresponding to the AV|Gated,
AV|Max, ParaFlow, Peak Combo sensor or Surface Combo sensor.
Note: Contact your regional ADS representative for
further diagnosis before replacing a sensor.
Problem
Pressure depth readings are erratic.
Possible
Causes
Pressure vent tube may be pinched or blocked with
water.
Pressure dryer tube may not be functioning
properly.
Black cap may still be on the dryer tube.
Desiccant has absorbed too much moisture and no
longer allows the dryer tube to function properly.
Possible
Solutions
Inspect color of desiccant in dryer tube and replace the
dryer tube if
necessary.
Inspect the vent tube along its entire length to ensure
the cable is not pinched and that the cable ties securing
the vent tube are not too tight.
Verify that the black cap has been removed from the
dryer tube.