Maintenance and Troubleshooting 11-69
Problem
Pressure depth readings display a good pattern, but are
consistently incorrect.
Possible
Cause
Sensor physical offset may be incorrect.
Possible
Solution
Verify the sensor offset and reactivate with the correct
settings if necessary.
Problem
Individual
PDEPTH_1 (_2)
values consistently read 0
under wet pipe conditions.
Possible
Causes
Sensor may be faulty.
5
-volt (blue) fuse on regulator may have blown.
Possible
Solutions
Replace the Peak Combo Sensor.
Replace the 5
-volt (blue) fuse on the regulator.
Caution:
Do not connect a questionable sensor to
another monitor before verifying the integrity of the
sensor. A shorted sensor will blow the 5
-volt (blue)
replacement fuse on the regulator in the monitor.
Temperature Subsystem
The following table contains general troubleshooting techniques for
the temperature subsystem corresponding to all sensors supporting
the TRITON+.
Note: Contact your ADS representative for further
diagnosis prior to replacing a sensor or monitor.