11-66 ADS TRITON+ Manual
Individual PEAKVEL_1 or PEAKVEL_3 values
consistently read
-9.99.
faulty.
-volt (blue) fuse on regulator may have blown.
Replace the Peak Combo Sensor or Surface Combo
Sensor.
-volt blue fuse on the regulator.
Do not connect a questionable sensor to
another monitor before verifying the integrity of the
sensor. A shorted sensor will blow the 5
-volt (blue)
replacement fuse on the regulator in the monitor.
Surface Velocity Subsystem
The following tables contain general troubleshooting techniques for
the surface velocity subsystem corresponding to the ParaFlow and
Surface Combo sensor.
Velocity readings are erratic or incorrect.
Sensor may be dirty or covered by
debris.
setting may require adjustment.
setting may require adjustment.
Clean the sensor transmit/receive surfaces.
-tune Maximum Velocity setting in device
Transmit Frequency
settings in the velocity
parameters.