11-62 ADS TRITON+ Manual
Problem
Individual UpDepth_1
(_2) values consistently read -
9.99.
Possible
Causes
Sensor may be faulty.
5
-volt (blue) fuse on regulator may be blown.
Possible
Solutions
Replace the Peak
Combo Sensor.
Replace the 5
-volt blue fuse on the regulator.
Caution:
Do not connect a questionable sensor to
another monitor before verifying the integrity of the
sensor. A shorted sensor will blow the 5
-volt (blue)
replacement fuse on the regulator in
the monitor.