22
Micro-Ohmmeter Model 6250
4.2.6 Setting the Language used for Printing Reports (LAnG)
• Turn the rotary switch to the SET-UP position.
• Press the ▲ button until “LAnG” appears on the top line of the
display.
• Press the ► button to enter the language selection sub-menu.
• Next, press the ▲ button to toggle between English (Lg Gb) or
French (Lg F).
• Press the ► button to validate the selection and return to the top
level of the Language set-up menu.
•
To proceed to the next programming variable, press the ▲ button.
4.2.7 Setting the Value for the Reference Temperature (trEF)
• Turn the rotary switch to the SET-UP position.
•
Press the ▲ button until “trEF” appears on the top line of the display.
• Press the ► button to enter the reference temperature sub-menu.
The current reference temperature will be displayed with the lead-
ing digit ashing.
• Use the ▲ button to change the leading digit. The minus (-) symbol
can also be programmed in this location.
• When the desired value is reached, press the ► button to move
the next digit to the right.
• Use the ▲ button to change this digit, then press the ► button to
again move to the next digit to the right.
• Repeat this process for each of the 5 digit locations.
•
After the 5th digit is programmed, press the ► button again to return
to the top level of the Reference Temperature set-up menu.
•
To proceed to the next programming variable, press the ▲ button.
NOTE: The program limits for the reference temperature are 32.0 to
130.0°F and -10.0 to 130.0°C. Attempting to set values outside these
limits will cause error message “Err23” (Entry Out of Range) to appear in
the display.
w ww . . co m