Micro-Ohmmeter Model 6250
5
CHAPTER 1
INTRODUCTION
WARNING
These safety warnings are provided to ensure the safety of person-
nel and proper operation of the instrument.
Do not attempt to perform any tests with this instrument until you
have read the user manual.
Tests are to be carried out on de-energized circuits only! Never
connect the unit to a live circuit.
When the unit is measuring resistance showing a high inductive
component (transformers, motors, etc.) after ending the measure-
ment, the unit then discharges the inductive sample and the
warning icon
appears for the entire duration. Never disconnect
the connection wires before this icon disappears.
The micro-ohmmeter should never be used in an explosive
environment (this includes poorly ventilated battery rooms and
enclosures).
Make sure the internal battery is fully charged prior to testing. If the
instrument has been left unused for several months, recharge the
battery.
We recommend recharging the micro-ohmmeter every month to
ensure a full battery charge when used.
When replacing fuses, install only fuses which are direct replace-
ments.
If the case needs cleaning, do not use any alcohol or oil based
cleaners. Preferably use soapy water with a damp cloth or
sponge.
The test leads and measuring wires must be in good condition and
should be changed if there is any evidence of deterioration (insula-
tion split, burnt, etc.).
Never exceed the safety values indicated in the specications.