Micro-Ohmmeter Model 6250
The Model 6250 then returns to the Stand-by state at the end of the mea-
surement. The unit is ready to perform another measurement.
M
O
C
M
O
C
OPER
MES
I
STBY
360ms
240ms
100ms
360ms
240ms
100ms
Figure 4-6
C = connection check
0 = residual voltage measurement
M = measurement of the voltage across the resistor terminals.
• Store the measurement by rst pressing the MEM button, then
select the object and test location to store the measurement using
the arrow buttons. When the desired location has been selected,
press the MEM button a second time to complete the data storage
process.
4.5.4 Low Inductive Resistance Automatic Triggering
Measurement Mode
This mode is intended only for measuring resistance on material without a
time constant (non-inductive). The use of the spring loaded Kelvin probes
(Cat. #2118.77 or Cat. #2118.78) listed in the accessories section are rec-
ommended for this mode.
From the Stand-by state, start a test by performing the following:
•
Press the button until the AUTO icon shows on the display.
• Press the START/STOP button.
• Connect the probes to the specimen. The unit waits until it detects
that the current and voltage leads are connected correctly.