40
Micro-Ohmmeter Model 6250
CHAPTER 5
MEMORY / PRINTING
5.1 Managing and Printing the Data in Memory
The memory is organized into locations called Objects and Tests.
There are 99 objects, each containing a maximum of 99 tests (measure-
ments). The maximum number of measurements that can be stored is
approximately 1500.
After taking a measurement, it may be stored in memory by pressing the
MEM button. The MEM icon blinks and the rst available Object and Test
location for storing this measurement is displayed with the last digit of the
test number blinking.
Example: Object 1 and Test 4 are displayed and the 4 will be blinking.
01:04.
OBJ : TEST
The object and test location can be changed using the arrow buttons.
• FREE is displayed when the location is empty
• OCC is displayed when the location has data stored in it
After choosing the measurement’s memory location, pressing the MEM
button a second time validates memory storage in a FREE location.
If an occupied location is selected, the OCC message blinks to warn
that this memory location is already taken. Storage action in this location
requires pressing the MEM button again. The previous measurement in
this location will be replaced by the new measurement.