Micro-Ohmmeter Model 6250
25
Press the ► button to enter the degree units sub-menu.
Use the
button
to toggle through the choices of Fahrenheit
(dEgF) or Centigrade (dEgC).
Press the button to validate the selection and return to the top
level of the Temperature units set-up menu.
To proceed to the next programming variable, press the ▲ button.
4.2.12 Setting Alarm Set Point, Direction and Buzzer Level (ALAr)
Turn the rotary switch to the SET-UP position.
Press the
button
until ALAr appears on the top line of the
display.
Press the button, Alarm 1 and its set point, direction and buzzer
level will appear with ALARM 1 ashing.
To modify the settings of Alarm 1, press the button. The direc-
tion arrow will now be ashing. To modify Alarm 2 press the
button
, ALARM 2 will then be ashing.
When the direction arrow is ashing, it may be changed by press-
ing the
▲ button
to toggle between HIGH (
activates above set
point) and LOW (
activates below set point).
Next, press the ► button to adjust the buzzer level for this alarm.
The buzzer icon will be ashing. The choices are OFF (no
icon displayed), LOW (small buzzer icon displayed ) or HIGH
(large buzzer icon
displayed). The selection is made by
pressing the
▲ button
while the icon is ashing.
Next, program the set point value used to trigger the alarm (ignor-
ing the decimal point) by pressing the ► button. The leading digit
will be ashing. Adjust the digit value using the
button
then
press the ► button to move to the next digit to the right.
Repeat the process for each of the ve digits.
Press the button when the numeric value is set to the desired
number. The decimal point and units (mΩ or Ω) will be ashing.
Set the desired resolution and units by pressing the
button
.
Each press will move the decimal point one place to the right.
Cycling through milliohms and then through ohms.
To accept the alarm settings once the decimal point and units
are at the desired values, press the button. This will bring you
to Alarm 2. Repeat the process as necessary to set Alarm 2’s
conditions.
w ww . . co m
information@itm.com1.800.561.8187

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AEMC 6250 and is the answer not in the manual?

AEMC 6250 Specifications

General IconGeneral
BrandAEMC
Model6250
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals