Purification Solution - Developer's Guide 21
Preparing Default Purification Methods
2
Preparative Method Settings
Fraction Collector (G1364B and its cluster unit)
In the module Configuration, verify that the preparative pump is linked.
Fraction Collector (G1364E, G7159B, G7166A and cluster unit)
In the module Configuration, verify that the preparative pump is linked.
Flow Modulator (G7170B)
NOTE
If the active splitter is activated by a contact other than contact A, use this contact
instead.
Fraction Trigger Mode Peak Based, max. peak duration: 1 min
Peak Detector G1390B (UIB II)
Mode: Off (if no auxiliary detector is connected to UIB II box)
Mode: On (if auxiliary detector is connected to UIB II box)
UV detector
Mode: Set one of modes and related values
Use MSD for mass-based Fraction Collection
Mark this check box for mass-based collection.
Fraction is collected when a peak is detected by
Select all peak detectors
Collection Behavior Select Enable Fraction Collection.
Peak Triggers Use
Mark the check box
Peak Detection Mode
Select a peak detection mode and set the related values.
Use MSD for mass-based Fraction Collection
Mark this check box for mass-based collection.
Trigger Combinations Select AND.
Splitting Operation Select Splitting enabled.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals