EasyManua.ls Logo

Agilent Technologies 2200 - Page 16

Agilent Technologies 2200
90 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16 Agilent 2200 TapeStation System - User Manual
2 Specifications
ScreenTape Specifications
Specification (D5000 ScreenTape Assay)
Analytical Specifications D5000 ScreenTape assay and reagents
Sizing Range 100 – 5000 bp
Typical Resolution 400 – 5000 bp: 15 %
Sensitivity
1
1
signal-to-noise >3 (single peak)
0.1 ng/µL
Sizing Precision
2
2
Determined using the D5000 Ladder as sample
5% CV
Sizing Accuracy
2,3
3
Accuracy for run using single ScreenTape device
±10 %
Quantification Precision 0.1 – 1 ng/µL: 15 % CV
1 – 50 ng/µL: 10 % CV
Quantification Accuracy
4
4
Measured against 2100 Bioanalyzer system
±20 %
Quantification Range 0.1 – 50 ng/µL
Maximum Sample Buffer Strength 250 mM KCl,
250 mM Tris-HCl,
125 mM NaCl,
50 mM Acetate,
25 mM MgCl
2
,
25 mM BSA,
25 mM Guanidine-HCl
Physical Specifications
Multiple ScreenTape (96 well)
compatibility
Not compatible
Analysis Time 16 samples: <25 min
Samples per consumable 1 ladder + 15 samples
Sample volume required 1 µL
Kit stability 4 months
Kit size 112 samples/box

Table of Contents