18 Agilent 2200 TapeStation System - User Manual
2 Specifications
ScreenTape Specifications
Specification (Genomic DNA ScreenTape Assay)
Analytical Specification Genomic DNA ScreenTape assay and reagents
Sizing Range 200 bp to > 60000 bp
Sensitivity 0.5 ng/µL
Sizing Precision
1
1
Determined using the Genomic DNA Ladder as sample
200 – 15000 bp: 15 % CV
Sizing Accuracy
1
200 – 15000 bp: ±15 %
Quantitative Precision
2
2
Average result from various genomic DNA sample types
15 % CV
Quantitative Accuracy
3
±20 %
Linear Concentration Range 10 – 100 ng/µL
DIN functional range 5 – 300 ng/µL
Physical Specification
Analysis Time 16 samples: < 25 min
96 samples: < 150 min
Samples per consumable 1 ladder + 15 samples
Sample Volume Required 1 µL
Shelf Life 4 months
Box/Kit size 112 samples/box