Agilent 2200 TapeStation System - User Manual 39
Using the 2200 TapeStation System
4
Operating Procedure
3 Place loading tips into the loading tip holder as shown and insert into
the 2200 TapeStation instrument.
4 Remove ScreenTape device from the foil packet.
NOTE
If any used tips are left in the tip-buckets, a pop-up window will ask for the discarded tips to
be removed. The 2200 TapeStation instrument will not run until all the tip buckets are
empty.
NOTE
A Loading tip transfer tool (G2964-60000) is available.
NOTE
Ensure that all 16 loading tips are inserted into the tip holder.
The laptop utilized for performing any previous use(s) of the ScreenTape device must be
utilized for all further re-use.
CAUTION
Damage to the 2200 TapeStation instrument and impact on performance
Failure to use the correct consumable components can cause damage to the
instrument.
Use the recommended loading tips.
Use the recommended foil plate seal.