EasyManua.ls Logo

Agilent Technologies 2200 - Page 74

Agilent Technologies 2200
90 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
74 Agilent 2200 TapeStation System - User Manual
5 Maintenance
Changing the Needle
Needle change intervals:
After 7680 lanes (3840 pierces in a Single loading system), the 2200
TapeStation Controller Software will inform the user that a needle
change is pending. The word Needle will appear in the bottom of the
software inside a yellow box.
After 8320 lanes (4160 pierces in a Single loading system), a needle
change is recommended. The box around the word Needle will change
from yellow to red.
After 8960 lanes (4480 pierces in a Single loading system), the needle
has completed its lifetime and must be changed before the 2200
TapeStation instrument will start.
Figure 7 Agilent 2200 TapeStation Controller Software indicating a Needle change is
recommended
Parts required # p/n Description
1 G2960-60062 Needle cartridge (for use in single pump systems)
For use with product numbers ST007, ST008 and ST009
OR 1 G2960-60063 Needle cartridge (for use in dual pump systems)
For use with product numbers ST017, ST019, ST010, G2960A,
G2961A, G2964AA, G2965AA and G2966AA

Table of Contents

Related product manuals