4349B
INITiate
Subsystem
INITiate Subsystem
The
INITiate
subsystem
controls
the initiation
of the
trigger system.
It indicates
all trigger
sequences
as
a
group
.The
detailed description
of trigger
system is
in \Trigger
System".
COMMAND
PARAMETER
INITiate
:CONTinuous
<
Boolean
>
[:IMMediate]
:INITiate[:IMMediate]
The
:INITiate[:IMMediate]
command
causes
all
sequences
to
exit
the
IDLE
state
and
enter
the
INITIA
TE
state
.
This
command
causes
the
trigger
system
to
initiate
and
complete
one full
trigger
cycle
,
returning
to
IDLE
on
completion.
(No
query)
If
the
4349B
is
not
in
IDLE
or
if
INITiate:CONTinuous
is
set to
ON
,
an
:INITiate:IMMediate
command
has
no
eect
on
the
trigger
system
and
an
error
0
213
is generated.
:INITiate:CONTinuous
f
ON
j
OFF
j
1
j
0
g
The
INITiate:CONTinuous
command
sets
or
queries
whether
or
not
the
trigger
system
is
being
continuously
initiated.
ON
,
1
continuously
initiated.
OFF
,
0
not
continuously
initiated.
Query
response
is
{1|0}
.
GPIB Reference 5-19

Table of Contents

Other manuals for Agilent Technologies 4349B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4349B and is the answer not in the manual?

Agilent Technologies 4349B Specifications

General IconGeneral
BrandAgilent Technologies
Model4349B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals