Chapter 5: Troubleshooting
To Configure the motherboard jumpers and setup BIOS
5–20
Figure 5-7
MO
T VP-22 Motherboard
Configure the MOT series VP-22 Motherboard BIOS parameters.
Use this procedure to set the MOT VP-22 motherboard BIOS.
1 Connect the power cable to the Infiniium oscilloscope.
2 Connect the external keyboard to the rear panel.
3 Press the delete key when you see the following prompt on the bottom of the screen
Press TAB to Show POST screen, DEL to enter SETUP, F12 to select boot device.
Note: If you are not seeing the prompt, or the oscilloscope does not appear to be functioning,
check the motherboard jumper setting, and the ribbon cable connectors. Otherwise, continue
with the next step.