Chapter 5: Troubleshooting
To Configure the motherboard jumpers and setup BIOS
5–21
BIO
S Setup Procedure
1
Go to Load Setting 2 Defaults and press Enter key. Select Y to load the defaults of
BIOS Setting 2 and press the Enter key.
2 If you have a power board part number 54855-66502 then perform the following steps.
Otherwise, go to step 7.
3 Go to Power Management Setup and press the Enter key.
4 Select the PWORN After PWR-Fail setting and press the Enter key.
5 Select the Off option and press the Enter key.
6 Return to the CMOS Setup Utility by pressing the Esc key.
7 Press F10 to save and exit the setup. Type “Y” to save changes.