Chapter 1: General Information
Specifications & characteristics
1–11
Stat
e Select one channel as clock, specify
other inputs as High, Low or X. Logic
Type: AND or NAND Setup time is 1 ns
and hold time is 0 ns.
Delay by Time Time: 5 ns to 10 ns. The trigger is
qualified by an edge. After the delay, a
rising/falling edge on any one selected
input will generate the trigger.
Delay by Events Events: 1 to 16,000,000 rising or falling
edges. The trigger is qualified by an
edge. After the delay, a rising/falling
edge on any one selected input will
generate the trigger.
Video Triggering 525 lines/60 Hz (NTSC), 625 lines/50 Hz
(PAL). Trigger on any line of Field 1 or
Field 2. User defined triggering: User can
specify sync pulse level, width and
polarity, edge number.
Violation Trigger
Setup/Hold
Modes: Setup, Hold or Setup and Hold.
Select Clock, Thresholds, setup and/or
hold time.
Pulse Width Triggers on pulse width >t or, <t.
Triggers on pulse as narrow as 500 ps.
Transition (Slew Rate) Select Rise Time or Fall Time,
present > t or present < t, thresholds.
Triggers on edges as fast 800 ps.
Accuracy for Glitch, Pulse Width,
and time qualified pattern
1.5 ns to 20 ns:
20 ns to 160 ms
±(20% of setting + 500 ps)
±(3% of setting + 2 ns)

Table of Contents

Other manuals for Agilent Technologies 54830 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 54830 Series and is the answer not in the manual?

Agilent Technologies 54830 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model54830 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals