Operation
Agilent 55B AA Spectrometer User’s Guide 85
Figure 39. Nitrous oxide/Acetylene flames. Left to right Lean, Stoichiometric, Rich
Performance checks
Check the performance of your nebulizer:
At installation.
Weekly.
When using a new type of solution.
After the nebulizer has been pulled apart and re-assembled.
To check the performance of your nebulizer:
1 Ensure the nebulizer is correctly installed and set up. Set the nebulizer to high
vacuum mode as per High vacuum settingon Page 82, or if you are using
organics, set the nebulizer to maximum uptake rate (thimble fully clockwise).
2 Prepare an aqueous solution of copper with a concentration of 5 mg/mL.
3 Set the instrument up to measure the absorbance of copper in an air-
acetylene flame.
4 With the system fully optimized as described in Optimizationon Page 74,
aspirate distilled water from a 10 mL calibrated cylinder for a carefully timed
period and calculate the uptake rate. This should be between 3 and 6.5
mL/min for a High Vacuum setting.
5 Zero the instrument (press the Alt and Read keys together).
6 Aspirate a solution of 5 mg/mL copper and note the absorbance reading.
Agilent guarantee an absorbance of 0.75 or better using the Mark 7 flame
atomization system under fully optimized conditions.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 55B AA and is the answer not in the manual?

Agilent Technologies 55B AA Specifications

General IconGeneral
BrandAgilent Technologies
Model55B AA
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals