Troubleshooting/Errors
142 Agilent 55B AA Spectrometer User’s Guide
9160 EEPROM checksum mono correction
The monochromator wavelength may not be accurate. Because of this the
instrument may fail to find a particular Hollow Cathode lamp Resonance Line,
resulting in
No Peak Detected errors. If the error persists, a service call may be
required.
9307 RBC/Mains frequency below 48Hz.
The instrument detected a power mains supply fault: Low mains frequency.
Possible causes include:
Power supply failure.
Faulty RBC (Rotating Beam Combiner) on double beam instruments only.
Faulty Mains Supply to the instrument.
If the error persists, a service call may be required.
9308 RBC/Mains frequency
The instrument detected a power mains supply fault: Mains frequency out of
range.
Possible causes include:
Power supply failure.
Faulty RBC (Rotating Beam Combiner) on double beam instruments only.
Faulty Mains Supply to the instrument.
If the error persists, a service call may be required.
9309 RBC/Mains frequency above 62Hz
The instrument detected a power mains supply fault: High mains frequency.
Possible causes include:
Power supply failure.
Faulty RBC (Rotating Beam Combiner) on double beam instruments only.
Faulty Mains Supply to the instrument.
If the error persists, a service call may be required.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 55B AA and is the answer not in the manual?

Agilent Technologies 55B AA Specifications

General IconGeneral
BrandAgilent Technologies
Model55B AA
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals