Troubleshooting/Errors
Agilent 55B AA Spectrometer User’s Guide 143
9310 Optical RBC frequency
Instrument fault. The instrument detected a problem with the RBC. A service call
may be required.
Optical sensor or associated circuitry may be faulty.
9311 Instrument fault +12V PSU
Instrument Fault. The instrument detected a problem with the +12 volt supply. A
service call may be required.
9312 Instrument fault -12V PSU
Instrument Fault. The instrument detected a problem with the -12 volt supply. A
service call may be required.
9313 Instrument fault 5V PSU
Instrument Fault. The instrument detected a problem with the 5 volt supply. A
service call may be required.
9316 Wavelength out of range
The Instrument is reporting that the current method has specified a wavelength
that is out of range for the Photomultiplier Tube (PMT).
Possible causes include:
A method wavelength suitable for a Wide Range PMT was specified but the
instrument is currently fitted with a Short Range PMT. Use a valid wavelength.
Currently available PMTs:
Wide range: up to 900 nm.
9317 No peak: low HC lamp energy
Low hollow cathode lamp energy.
The instrument Gain had been raised to its highest value but the hollow cathode
lamp signal was too low.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 55B AA and is the answer not in the manual?

Agilent Technologies 55B AA Specifications

General IconGeneral
BrandAgilent Technologies
Model55B AA
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals